Timer Calibration Verification - Guidelines
Timer Calibration VerificationGuidelineNumber
Equ15-01-G
V.3.0
Effective Date 25 Sept 08
Subject
Timer Calibration Verification
Page 1 of 1
Supersedes Equ15-01-GV.2.0
Review by Heidi Hanes Review date 9-Feb-12
SMILE Comments: This document is provided as an example only. It must be revised toaccurately reflect your lab’s specific processes and/or specific protocol requirements.Users are directed to countercheck facts when considering their use in other applications.If you have any questions contact SMILE.
Audi t: Equipment Section
CAP Accreditation Checklis t:
1. Limited service laboratory checklist manual, Instrument andEquipment
2. Hematology –Coagulation checklist Manual, Instrument andEquipment
3. Chemistry and Toxicology checklist manual, Methods andInstrument systems
Background Information:
1 Timer calibration is to be verified twice per year after maintenance
8/13/2019 Equ15-01 Timer Cal Verfn SOP-App
http://slidepdf.com/reader/full/equ15-01-timer-cal-verfn-sop-app 2/14
SMILEJohns Hopkins University
Baltimore, MD USA
Timer Calibration Verification - Policy
Timer Calibration Verification PolicyPolicy Number Equ15-01-P V.3.0
Effective Date 25 Sept. 08
Approval Signature: Laboratory Director Last Review 27 July 2010
Supersedes Equ15-01-P V.2.0
Policy The laboratory has processes and procedures to effectively ensurethat timers are maintained and calibrated to ensure that theiraccuracy and precision is preserved.
Purpose This policy provides direction for the processes and procedures toeffectively ensure that the maintenance and calibration of timers iscarried out as required to ensure the accuracy and precision of theinstruments.
Responsibility The Laboratory Medical Director is responsible for ( to be determinedby Laboratory )
The Laboratory Manager is responsible for ( to be determined by theLaboratory )
The Lead Technologists are responsible for ( to be determined by
the Laboratory ) AdditionalInformation as
N/A
8/13/2019 Equ15-01 Timer Cal Verfn SOP-App
http://slidepdf.com/reader/full/equ15-01-timer-cal-verfn-sop-app 3/14
SMILEJohns Hopkins University
Baltimore, MD USA
Timer Calibration Verifi cation - Process
Timer Calibration VerificationProcess
Policy Number Equ15-01-Pro V3.0
Effective Date 25 Sept. 08
Approval Signature: LaboratoryDirector
Last Review 26-Oct-2011
Supersedes N/A Equ15-01-Pro V2.0
Process for Timer Calibration Verification
What Happens Who’s Responsible Instruct ions
Maintenance andCalibration policy forTimers is determined
• To be determined byLaboratory
Technical SOPs thatrequire the use of a
timer
Timer Maintenance andCalibration is performed
• QA Technologist• Technicians/Technologists
Timer Maintenanceand Calibration SOP
8/13/2019 Equ15-01 Timer Cal Verfn SOP-App
http://slidepdf.com/reader/full/equ15-01-timer-cal-verfn-sop-app 4/14
SMILEJohns Hopkins University
Baltimore, MD USA
Author(s), Name &Title
Kurt Michael M.ED., MT (ASCP)
Josephine Shim MBA, MT(ASCP)
DocumentNumber
EffectiveDate
QA/QC International ManagerEqu15-01- SOP
V.4.0
25 Sept. 08
ApprovedBy
Name, Title Signature Date
SOP AnnualReview
Name, Title Signature Date
RevisionHistory
Version #[1.0]
Revision Date[dd/mm/yy] Description (notes)
Sections A and B added to the
8/13/2019 Equ15-01 Timer Cal Verfn SOP-App
http://slidepdf.com/reader/full/equ15-01-timer-cal-verfn-sop-app 5/14
SMILEJohns Hopkins University
Baltimore, MD USA
I acknowledge that I have read, understand and agree to follow this SOP.
Name (print) Signature Date
8/13/2019 Equ15-01 Timer Cal Verfn SOP-App
http://slidepdf.com/reader/full/equ15-01-timer-cal-verfn-sop-app 6/14
SMILEJohns Hopkins University
Baltimore, MD USA
Timer Calibration Verification - SOP
Purpose
1. To ensure timers used in procedures are accurate when comparedto a current certified traceable timer/stopwatch such as a NationalInstitute of Standards and Technology (NIST) traceable stopwatch.
Definition
1. NIST Certified timer: a timer that has been verified as accurate froman outside, NIST source.
2. NIST verified timer: a timer that has been verified as accurateagainst a NIST in house source (such as a website). This can beaccomplished by following either procedure A below.
Pre-analytic Procedure1. This procedure applies to all laboratory timers including those timers
that are included as parts of other instruments such as centrifuges.2. Timers that are not verified using the procedure outlined in this SOP
may be cleaned or serviced and retested. If a timer cannot berepaired it should be discarded.
3. Calibration records should indicate dates of removal, return to
service or if timer is discarded.4. All calibration sheets should remain on file as a record, even afterthe timer has been removed permanently from service.
5. At time of calibration verification, timers should be inspected for
8/13/2019 Equ15-01 Timer Cal Verfn SOP-App
http://slidepdf.com/reader/full/equ15-01-timer-cal-verfn-sop-app 7/14
SMILEJohns Hopkins University
Baltimore, MD USA
Timer Calibration Procedure
If using an Uncertified Timer, the timer will need to be verified first beforetesting other timers in the laboratory. Verify the timer by following theinstructions outlined in either Section A and then use the verified timer toperform timer checks as outlined in Section B.
Note : Timers used at the bench are to be considered contaminated andshould only be handled with gloves and other personal protectiveequipment and/or thoroughly disinfected before calibration verification.
A. Procedure for Verification of a laboratory Timer against NIST TraceableWebsite
1. Locate the official National Institute of Standards and Technology atNIST traceable website at http://www.nist.gov/index.html . Scrolldown to the bottom of the page to see the NIST clock on the right
hand side.
8/13/2019 Equ15-01 Timer Cal Verfn SOP-App
http://slidepdf.com/reader/full/equ15-01-timer-cal-verfn-sop-app 8/14
SMILEJohns Hopkins University
Baltimore, MD USA
2. If desired, you can follow the instructions on the site to post the“Official U.S Time” widget to your computer..
3. Please note that it is the time interval, not the actual time that isimportant for the timer verification process.
4. Once you have located the NIST timer you can change the timezone using the arrows to the left and right.
5. Take the timer to be checked to the same location as the computerused to access the NIST website.
6. At the beginning of the hour start the laboratory timer that is beingverified Document the time for both the computer website and the
8/13/2019 Equ15-01 Timer Cal Verfn SOP-App
http://slidepdf.com/reader/full/equ15-01-timer-cal-verfn-sop-app 9/14
SMILEJohns Hopkins University
Baltimore, MD USA
Tolerance Limits for Verification of a Timer against NIST Traceable Source
The NIST verified timer should read one hour (60 Minutes) +/- 1 secondat the end of the test time of one hour.
Once the timer has gone through the verification procedure and haspassed, this timer can then be used to check other timers in thelaboratory .
Note: The tolerance limits for the NIST verified timer are different fromthe limits listed under section B. The limits for the NIST verified timerare more stringent since it will be used as the reference method.
B. Equipment Calibration/Maintenance:
1. Timers used at the bench are to be considered as contaminated andshould only be handled with gloves and other personal protectiveequipment and/or thoroughly disinfected before calibrationverification.
2. Set the laboratory timer at a setting used in a procedure, and startthe Test timer and the NIST certified or verified timersimultaneously.
3. Stop the NIST verified timer precisely as the timer being checkedindicates that the set time has elapsed.
4. Record times from both the verified timer and the “Test Timer” onthe Timer Calibration Verification Record Sheet (Appendix B).
5. Calculate the difference between the two times and record. If there
8/13/2019 Equ15-01 Timer Cal Verfn SOP-App
http://slidepdf.com/reader/full/equ15-01-timer-cal-verfn-sop-app 10/14
SMILEJohns Hopkins University
Baltimore, MD USA
10. Repeat the process for all times called for in the testing proceduresfor which this timer is used.
Quality Control: N/A
Test Method Instructions: N/A
Calculations: N/A
Tolerance Limits for Timer Verification:
Acceptable difference between the “Test Timer” and the Certified orVerified timer must be ± 2% of the total “Test Timer” setting or as themanufacturer recommends and in accordance with testing procedure.
Method Limitations: N/A
Post-analytic Procedure
How are Results Prioritized: N/A
Notification of Result to Appropriate Personnel: N/A
8/13/2019 Equ15-01 Timer Cal Verfn SOP-App
http://slidepdf.com/reader/full/equ15-01-timer-cal-verfn-sop-app 11/14
SMILEJohns Hopkins University
Baltimore, MD USA
3. Joint Commission on Accreditation of Healthcare Organizations,How to Meet the Most Frequently Cited Laboratory Standards, 1996,Oakbrook Terrace, IL.
4. Timer Verification SOP V1.0 , Muhimbili University of Health and Allied Sciences, Dar Es Salaam, Tanzania.
5. JHU Immunogenetics Laboratory : Digital Timer SOP and Log, JHUSchool of Medicine, November 2000 Baltimore, MD, USA.
Appendices
1. Appendix A -Timer Verification for Non-Certified Timer against NISTTraceable Source - Record Sheet
2. Appendix B – Example of Timer Calibration Verification RecordSheet
3. Appendix C - NIST Calibration Tool Source (attached)
8/13/2019 Equ15-01 Timer Cal Verfn SOP-App
http://slidepdf.com/reader/full/equ15-01-timer-cal-verfn-sop-app 12/14
SMILEJohns Hopkins University
Baltimore, MD USA
APPENDIX C
NIST Calibration Tool Source
NIST instruments can be purchased from Fisher Scientific athttps://www1.fishersci.com/ .
NIST thermometer: Catalog# 15-077-55
NIST timer: Catalog# 14-649-11
NIST tachometer: Catalog# 05-028-23
Shipping will be extra. They will charge a hazardous fee for the tachometerbattery, but it may be negotiable.
8/13/2019 Equ15-01 Timer Cal Verfn SOP-App
http://slidepdf.com/reader/full/equ15-01-timer-cal-verfn-sop-app 13/14
Equ1.5-01 Appendix A
Timer Acceptability Criteria: The NIST verified timer should read one hour +/- 1 second at the end of the test time of one hour.
Reviewed by: ____________________ Date: ___________________________________
Reviewer's Comments: ____________________________________________________________
Date
App endix A: Timer Ver if icati on fo r Non-Cer ti fi ed Timer agains t NIST Traceabl e Sou rce - Record Sheet
St ar t t ime Stop t ime To tal t ime St ar t t ime
(Laboratory Name)
Difference -Total Time Aminus Total
Time B
Acceptable/Unacceptable
TechInitials
Stop time Total t ime
NIST Source (A) Non Certified Timer (B)
LaboratoryNon-Certified
Timer ID#
TimeInterval forVerification
8/13/2019 Equ15-01 Timer Cal Verfn SOP-App
http://slidepdf.com/reader/full/equ15-01-timer-cal-verfn-sop-app 14/14
Appendex I - Timer Cal ibrat ion Verif ication Record Sheet
Equ15-01_App_B_Timer_Cal_Verfn_Sht.xls Version cal/ver101-AP1
Timer # :___________ Locati on of timer:_______________________ *one timer per sheet
Timer Criteria: Acceptable difference must be equal to or less than 2% of the timer setting.
Reviewed by:__________________________________ Date:_____________________
Appendix B - Timer Calibration Veri fi cat ion Record Sheet
DateTech
(Initials)
Time Intervalfor
Verification
NIST Calibr atedTimer - Actual
Time
TimeDifference
Timer - ActualTime
List testing procedures for which the timer is used:____________________
Accep tabl eRange
SupervisorReview (Intials
and Date)
Acceptab le/Unacceptable