Calibracion de Temperatura

download Calibracion de Temperatura

of 8

Transcript of Calibracion de Temperatura

  • 7/31/2019 Calibracion de Temperatura

    1/8

    ZERO SPAN

    4 to 20 mA

    4 Wire RTDTransmitter

    RTD Sensor

    2200 C

    0

    50

    100

    150

    200

    250300

    -200 -100 0 100 200 300 400

    Ohms Vs. Temp (PT100)

    Technical Data

    Temperature Calibration

    Fluke 724 Temperature Calibrator

    Function

    Fluke Temperature Test Tools

    Fluke

    712

    Fluke

    714

    Fluke

    724

    Fluke

    725

    Fluke

    741

    Fluke

    743

    Fluke

    744

    Fluke

    55XX

    Fluke

    50

    Series

    Temperature plays a key role inmany industrial and commercialprocesses. Examples includemonitoring cooking temperature

    in food processing, measuringthe temperature of molten steelin a mill, verifying the tempera-ture in a cold storage warehouseor refrigeration system, or regu-lating temperatures in the dryingrooms of a paper manufacturer.

    A temperature transmitter willuse a measuring device to sensethe temperature, and then regu-late a 4-20 mA feed-back loop to a controlelement that affectsthe temperature (Fig.1). The control ele-

    ment might consist ofa valve that opens orcloses to allow moresteam into a heatingprocess or more fuel

    to a burner. The two most com-mon types of temperature sens-ing devices are the thermocouple(TC) and resistive temperature

    detector (RTD).Fluke provides a broad rangeof temperature calibration toolsto help you quickly and reliablycalibrate your temperature in-strumentation. A summary of thetemperature calibration capabili-ties of Fluke Process Tools isshown below.

    Figure 1.

    Measure temperature from an RTD probe

    Measure temperature from a T/C probe Simulate an RTD output Simulate an RTD into pulsed excitation current Simulate a T/C output Simultaneous output a T/C, measure mA Simultaneous output an RTD, measure mA Log a temperature measurement Ramp a temperature signal Loop power supply Multifunction Source and Measure Automatically calibrate temperature switches Electronic data capture Upload documented data to PC Integrated HART communication

  • 7/31/2019 Calibracion de Temperatura

    2/8

    2 Fluke Corporation Temperature Calibration

    Typical Temperature Calibration Applications

    How to calibrate a Thermocouple input transmitter

    The Fluke 724 TemperatureCalibrator can provide the threethings necessary to calibrate atemperature transmitter. You cansource a temperature, provideloop power, and measure theresulting output current. Thefollowing example shows how tocalibrate a Type K TC transmitterthat is ranged from 0-150degrees centigrade, generatingan output current range from4-20 mA.

    Basic Calibrator Setup

    1. Connect the 724 test leads tothe TC transmitter as shown.The output from the thermo-couple jacks on the 724 willsimulate a temperature inputto the transmitter. The redand black test leads willprovide loop power to thetransmitter and will measurethe current resulting fromtemperature changes into thetransmitter.

    2. Power on the 724 calibrator.Select the mA button and theLOOP button to select mea-

    sure milliamps with 24V looppower applied.

    3. Press the Meas/Source buttonuntil the lower portion of the724 display indicates thesource mode.

    4. Depress the TC button until aTC type of K is displayed.

    5. Select the C button forcentigrade.

    6. Set the Zero Point for thisapplication into the Calibra-tor. To do this set the displayinitially to 0.0C. You can usethe up and down arrow keysto change the output value.Use the left and right arrowsto control which decadevalue of the display is beingchanged. When the displayreads 0.0, hold down the 0%key on the 724 and observethat 0% is displayed in thelower right corner of thescreen. This establishes theZero point for calibration.

    7. Set the Span Point in theCalibrator. Set the display tothe desired Span value forcalibration. In this examplethe display should read

    150C. Depress the 100% keyand observe that 100% isdisplayed in the lower rightcorner of the screen. Thisestablishes the Span point forcalibration.

    Performing an As Found Test8. Depress the 0% key; record

    the applied temperature andthe corresponding mAmeasurement.

    9. Depress the 25% key (2)times; record the appliedtemperature and the corre-sponding mA measurement.

    10.Depress the 100% key; recordthe applied temperature andthe corresponding mAmeasurement.

    11. Calculate the errors for eachof the (3) points using thefollowing formula: ERROR =([(I-4)/16]-[(T/TSPAN])*100where Error is in % of span, Iis your recorded mA mea-surement, T is the recordedtemperature and TSPAN is thetemperature input span(100% - 0% points). The errorcalculation table belowshows how to apply theformula to actual recordedmeasurements.

    12.If your calculated errors areless than the specifiedinstrument tolerance, the

    transmitter has passed theAs-Found test. If the test hasnot passed, perform adjust-ments as necessary.

    Adjusting the Transmitter13.Depress the 0% key to source

    the proper temperature for a4 mA output. Adjust the zeropotentiometer until thecurrent reading is 4.00 mA.

    14.Depress the 100% key tosource the proper temperaturefor a 20 mA output. Adjust theSpan potentiometer until thecurrent reading is 20.00 mA.

    15.Depress the 0% key againand adjust the zero potenti-ometer again if necessary, toget a 4.00 mA output.

    Perform an As Left TestRepeat steps 8 through 12 tocomplete the full calibrationprocedure on your temperaturetransmitter.

    mA Measurement TC Source T Span Formula Error %

    4.02 0C 150C ([4.02-4)/16]-[0/150])*100 0.1250

    11.95 75C 150C ([11.95-4)/16]-[75/150])*100 -0.3125

    20.25 150C 150C ([20.25-4)/16]-[150/150])*100 1.5625

    Temperature transmitter error calculation example

    100%

    25%

    25%

    RECALL

    MEAS

    SOURCE

    STORESETUP

    TC RTD

    0%

    TEMPERATURECALIBRATOR724

    V

    F

    C

    V LOOPmA

    VmA

    MEASURESOURCE/MEASURE

    30VMAX ALLTERMINALS

    4W

    3W V

    RTD

    TC

    COM COM

    LOOP

    TEST DC PWR++

    Rosemount 444

  • 7/31/2019 Calibracion de Temperatura

    3/8

  • 7/31/2019 Calibracion de Temperatura

    4/8

    4 Fluke Corporation Temperature Calibration

    Fluke 725 Multifunction Process Calibrator All the temperature capabilities of Fluke 724

    below, plus: Measure pressure with Fluke 700Pxx

    Pressure Modules Source or measure frequency to 10 kHz Source mA signals

    Fluke 741, 743 and 744 DocumentingProcess Calibrators

    Measure or simulate JKTERSBLUNCthermocouples

    Measure or simulate 7 RTD types, per 712below, plus Cu 10 (47)

    Source or measure volts, ohms, mA, frequency Measure pressure with Fluke 700Pxx

    Pressure Modules Electronically capture results of automated

    procedures Serial communication to PC (743, 744) HART communication (744)

    Fluke 724 TemperatureCalibrator

    Measure or simulate JKTERSBLUN thermocouples

    Measure or simulate 7 RTDtypes (see 712)

    Measure or source VDC, Measure 24 mA, with or

    without 24 volt loop supply Source TC or RTD while

    measuring V or mA Handles pulsed RTD transmit-

    ters with pulses >5 ms

    Fluke 712 RTD Calibrator

    Measure or simulatePt 100 200 500 1000 (385),Pt 100 (3926), Pt 100 (3916)and Ni 120 (672)

    Measure or source 15 to3200

    Handles pulsed RTDtransmitters

    Selectable F or C Banana jacks for 2W, 3W or

    4W measurements

    Fluke 714 ThermocoupleCalibrator

    Measure or simulate JKTERSBLU thermocouples

    Measure or source -10 to 75 mV Selectable F or C Thermocouple mini-jack

    termination

  • 7/31/2019 Calibracion de Temperatura

    5/8

    Temperature Calibration Fluke Corporation 5

    Fluke 51 Series II and 52 Series II

    Large backlit dual display shows any combina-tion of T1, T2 (52 only), T1-T2 (52 only), plusMIN, MAX, or AVG

    Relative time clock on MIN, MAX, and AVG pro-vides a time reference for major events.

    Electronic Offset function allows compensationof thermocouple errors to maximizeoverall accuracy

    Measures J, K, T, and E-types ofthermocouples

    Readout in C, F, or Kelvin (K).

    Fluke 51 Single InputDigital Thermometer

    Fluke 52 Dual InputDigital Thermometer

    Fluke 53 Single Input DigitalThermometer with data logging

    Fluke 54 Dual Input DigitalThermometer with data logging

    The Fluke 53 Series II and 54 Series II

    Include all the features listed for the 51and 52 Series II plus: Data Logging up to 500 points of data with user

    adjustable recording interval Additional thermocouple types R, S, and N (for a

    total of 7 different types) Real time clock captures the exact time of day

    when events occur Recall function allows logged data to be

    easily reviewed on the meter display IR communication port allows data to be ex-

    ported to optional FlukeView Temperature PCsoftware for further analysis and graphing

    65 Infrared Thermometer

    Range -40 to 500C (-40 to932F)

    0.1 resolution up to 200 Measurements in less than

    1 second Bright laser beam for easy

    targeting 8:1 Optical Resolution MIN/MAX reading capture Large backlit dual display

    2620T/2635T RecordingThermometers 20 channel precision

    thermometer with SPRTReference Probe

    Up to 18 channels TCCalibration System

    Up to 9 channels RTDCalibration System

    NIST/DKD certifiedcalibration to .05C

    61 Infrared Thermometer

    Easy to use Quick non-contact

    temperature measurementsfrom -18 to 275 C(0 to 525 F)

    0.2 C or 0.5 F resolution Backlight illuminates the

    display in the dark Laser-sighted; 8:1 optical

    resolution

  • 7/31/2019 Calibracion de Temperatura

    6/8

    6 Fluke Corporation Temperature Calibration

    Temperature Test Tool Summary Performance: Selected examples

    Function Range Resolution Accuracy Notes

    RTD Measure -200 to 800C 0.1C 0.3C 8 RTD typesPT100-385 Simulate -200 to 800C

    RTD Measure -200 to 800C 0.1C 0.33C 7 RTD typesPT100-385 Simulate -200 to 800C 0.1C 0.33C

    Resistance Measure 0-11,000 .01 .05% + 50 m100 range Simulate 0-11,000 .01 .01% + 40 m

    Resistance Measure 15-3200 .1 .1 to 1100 range Simulate 15-3200 .1 .1 to 1

    Thermocouple Measure -202 to 1374C 0.1C 0.3C JKTERSBLUNCType K Simulate -202 to 1374C 0.1C 0.3C

    Thermocouple Measure 0.1C 1000C + 0.3C

    Thermocouple Measure 0.1C 1000C + 0.3C

    Thermocouple Measure -200 to 1370C 0.1C 0.3C + 10 V JKTERSBLU plusType K Simulate -200 to 1370C 0.1C 0.3C + 10 V N on 724, 725

    mV Measure 0-110 mV .001 mV .25% + .015% FSSource 0-110 mV .001 mV .01% + .005% FS

    mV Measure -10 to 75 mV .01 mV .025% + 1 countSource -10 to 75 mV .01 mV .025% + 1 count

    Measure mA 0-24 mA .001 mA .020% + 2 count Measure mA 0-24 mA .001 mA .010% + .015% FS Loop Power Supply 24V dc n/a 10%

    Dry Well Calibrator: A tempera-ture calibrator that uses a

    precision oven to source precisetemperature. This style ofcalibrator is often used for theverification of temperaturesensors.

    Excitation current: A constantcurrent applied to an RTD probeto determine actual resistancefor temperature measurement.Typical values are 2 mA or lessto minimize self-heating of theprobe.

    IPTS-68: International Practical

    Temperature Scale of 1968. Atemperature standard adopted in1968 that uses intrinsic stan-dards to define the measurementof temperature.

    ITS-90: International Tempera-ture Scale of 1990. A tempera-

    ture standard adopted in 1990that uses intrinsic standards todefine the measurement oftemperature. This standardmodifies the intrinsic standardsof IPTS-68 with additionalintrinsic references.

    Lead Resistance Compensation:A compensation method usedwith 3 and 4 wire RTDs andresistance measurement. Thismethod negates the error associ-ated with lead resistance whenmaking an RTD measurement.

    Reference Temperature: Thetemperature at which a thermo-couple temperature measure-ment is referenced. For FlukeCalibration tools, this is thetemperature where the thermo-couple mini-plug is connected tothe calibrator.

    R: The resistance value of anRTD probe at 0C. Example

    PT100-385, R = 100.RTD: Resistance TemperatureDevice, a temperature measure-ment sensor that has predictablechanges in resistance with achange in temperature. The mostcommon RTD is the platinumPT100-385.

    Seebeck Effect: Thermoelectriceffect in which the voltagepotential increases with tem-perature (thermocouples) in ajunction of dissimilar metals.

    Triple point of water: Thistemperature reference point isthe intrinsic standard at whichwater is liquid, ice and gas. Thisreference point defines 0.01C.

    Temperature Terminology

    744

    743/743B

    741/741B

    725

    724

    714

    712

    53/54

    51/52

  • 7/31/2019 Calibracion de Temperatura

    7/8

    Temperature Calibration Fluke Corporation 7

    Alloys Fluke 700 Series DPCs

    Plug + - Temperature EMF RangeColor Range

    White Pt-30% Rh Pt-6% Rh 600 to 1820C, 1.792 to 13.820 mV

    B Platinum- Platinum- 1112 to 3308F30% Rhodium 6% RhodiumRed W-5% Re W-26% Re 0 to 2316C, 0 to 37.079 mV

    C Tungsten- Tungsten- 32 to 4201F5% Rhenium 26% RheniumPurple Ni-Cr Cu-Ni -250 to 1000C, -9.719 to 76.370 mV

    E Nickel- Copper-Nickel -418 to 1832FChromium (Constantan)(Chromel)

    Black Iron Fe Cu-Ni -210 to 1200C, -8.096 to 69.555 mV

    J Copper-Nickel -346 to 2193F(Constantan)Yellow Ni-Cr Ni-Al -200 to 1372C, -5.891 to 54.886 mV

    K Nickel- Nickel- -328 to 2502FChromium Aluminum(Chromel)Blue Fe Cu-Ni -200 to 900C, -8.166 to 53.147 mV

    L Iron Copper-Nickel -328 to 1652F(Constantan)Orange Ni-Cr-Si Ni-Si-Mg -200 to 1300C, -3.990 to 47.514 mV

    N Nickel-Chromium- Nickel-Silicon- -328 to 2372FSilicon Magnesium(Nicrosil) (Nisil)

    Green Pt-13% Rh Pt -20 to 1767C, -0.101 to 21.089 mV

    R Platinum- Platinum -4 to 3213F13% RhodiumGreen Pt-10% Rh Pt -20 to 1767C, -0.103 to 18.682 mV

    S

    Platinum- Platinum -4 to 3213F10% Rhodium

    Blue Cu Cu-Ni -250 to 400C, -6.181 to 20.873 mV

    T Copper Copper-Nickel -418 to 752F(Constantan)White Cu Cu-Ni -200 to 600C, -5.693 to 34.320 mV

    U Copper Copper-Nickel -328 to 1112F

    Thermocouples TemperatureConversion

    F

    4352

    4172

    3992

    3812

    3632

    3452

    3272

    3092

    2912

    2732

    2552

    2372

    2192

    2012

    1832

    1652

    1472

    1292

    1112

    932

    752

    572

    392

    212

    32

    -148

    -328

    C

    2400

    2300

    2200

    2100

    2000

    1900

    1800

    1700

    1600

    1500

    1400

    1300

    1200

    1100

    1000

    900

    800

    700

    600

    500

    400

    300

    200

    100

    0

    -100

    -200

    -459 -273

    Fixed Points of ITS-90

    961.78C Silver freezing point

    660.323C Al freezing point

    419.527C Zinc freezing point

    -38.8344C Mercury triple point

    231.928C Tin freezing point

    156.5985C Indium freezing point

    0.01C H2O triple point

    29.7646C Gallium melting point

    -189.3442C Argon triple point-218.7916C Oxygen triple point-248.5939C Neon triple point-259.3467C Hydrogen triple point

    Application AssistanceTo request copies of these application notes,contact your Fluke sales organizations.

    Understanding Specifications (pub ID 10127) Transmitter Calibration (pub ID 10123) HART Transmitter Calibration (pub ID 10179) Limit Switches (pub ID 10121) Pressure Modules (pub ID 10188) Loop Calibration (pub ID 10437)

  • 7/31/2019 Calibracion de Temperatura

    8/8

    Fluke 700TC1 ThermocoupleMini-Plug Kit (11 types)For use with: Fluke 700, 720 or

    740 Series Process Calibrators, Fluke714 Thermocouple Thermometer.Description: A kit of 10 mini-plugconnectors:Type J (black), oneType K (yellow), oneType T (blue), oneType E (purple), oneType R/S (green), oneType B or Cu (white), oneType L (J-DIN) (blue), oneType U (T-DIN) (brown), oneType C (red), oneType N (orange), one

    Fluke 700TC2 ThermocoupleMini-Plug Kit Type J, K, T, E, R/SFor use with: Fluke 700, 720 or 740Series Process Calibrators, Fluke 714Thermocouple Thermometer.Description: A kit of seven mini-plugconnectors:Type J (black), twoType K (yellow), twoType T (blue), oneType E (purple), oneType R/S (green), one

    80T-IR Non-Contact InfraredTemperature ProbesFor quick, non-contact temperaturemeasurement in less than onesecond Instantly measures

    temperatures from -18 C to260 C (0 to 500 F)

    Highly repeatable readingsdisplayed in C or F

    Optical resolution 4:1 Output 1mV/C or 1mV/F

    (switch selectable)

    80T-IR/E Extended Range,Non-Contact InfraredTemperature ProbeSpecial version of the 80T-IR:

    Extended range: 0 to 1,000 F Measures F only (no C)80TK Thermocouple ModuleConverts multimeters with a mVrange into digital thermometers. Converts K-type thermocouple

    signals into mV output. Connects to DMM via

    standard bannana plugs Switch selectable C or F Range: -50 C to 1000 C

    (-58 to 1832 F)

    Infrared Temperature Probe -80PK-IRCompatible with Fluke 714, 724,

    725, 74X, Fluke 50 Series thermom-eters, Fluke 16 and 89 IV DMMs, thisprobe measures the temperature ofobjects that cant be touched. Measurement range: -18 to

    260 C (0 to 500 F)

    General Purpose Bead Probes80PK-1 and 80PJ-1Basic, inexpensive thermocouple.Good accuracy and fast response.(Not suitable for liquid immersion.) 80PK-1 for K-type thermometers;

    80PJ-1 works for J-Typethermometers

    Measurement range: -40 to260 C (-40 to 500 F)K-Type Immersion Probe -80PK-2AGeneral-purpose probe includingliquids and gels. Measurement range: -40 to

    982 C (-40 to 1800 F)

    K-Type Flat Surface Probe -80PK-3AMeasure the temperature of flat orslightly convex surfaces, such as hotrollers and plates. Exposed junction for direct

    contact with surface Measurement range: 0 to 260 C(32 to 500 F)

    K-Type Air Probe80PK-4AFor air and non-caustic gasapplications. Probe of 304 stainless steel with

    perforated baffle around the bead Measurement range: -40 to 816C

    (-40 to 1500 F)

    Piercing Probes -80PK-5A and 80PT-5ADesigned for foods, liquids and gels.

    Probe made of 304 stainless steel 80PK-5A is compatible withK-Type instruments; 80PT-5Ais designed for use with T-Typethermometers

    Measurement range: -40 to260 C (-40 to 500 F)

    K-type Exposed Junction Probe -80PK-6AUse both as a surface probe and forinsertion into air and non-corrosivegases. Probe made of 304 stainless steel Measurement range: -40 to 816 C

    (-40 to 1500 F)

    K-Type Industrial / Heavy DutySurface - Probe - 80PK-7For flat or slightly curved surfaces.

    Exposed junction allows directcontact with the surface beingmeasured

    Rugged design for extendedservice life

    Measurement range: -127 to600 C (-196 F to 1112 F)

    K-Type Probe Pipe Clamp Tem-perature Probe-80PK-8Designed for pipes from 6.4 mm(1/4") diameter to 34.9 mm(1-3/8") diameter. Measurement range: -29 C to

    149C (-20 F to 300 F)

    General Purpose Probes -80PK-9 and 80PJ-9Use as surface probe and forinsertion into air and non-causticgases. Probe constructed of 304

    stainless steel 80PK-9 is compatible with K-type

    thermometers; 80PJ-9 operateswith J-Type

    Measurement range:-40 to 260 C (-40 to 500 F)

    Extension Wire Kits - 80PK-EXT,80PJ-EXT and 80PT-EXT

    For extending and repairing J, K orT-type thermocouple wires. Kit includes 3 meters of thermo-

    couple wire and 1 pair ofmale/female mini-connectors

    Maximum continuous exposuretemperature: 260 C (500 F)

    Temperature Accessories

    Fluke CorporationPO Box 9090, Everett, WA USA 98206

    Fluke Europe B.V.PO Box 1186, 5602 BDEindhoven, The Netherlands

    For more information call:U.S.A. (800) 443-5853 orFax (425) 446-5116Europe/M-East/Africa(31 40) 2 675 200 orFax (31 40) 2 675 222Canada (800) 36-FLUKE orFax (905) 890-6866

    Other countries +1 (425) 446-5500 orFax +1 (425) 446-5116Web access: http://www.fluke.com

    2003 Fluke Corporation. All rights reserved.Printed in the Netherlands 4/2003 Pub ID 10308-engRev 02

    Fluke. Keeping your worldup and running.