Technical Data
Temperature Calibration
Fluke 724 Temperature Calibrator
Function
Fluke Temperature Test Tools
Fluke
712
Fluke
714
Fluke
724
Fluke
725
Fluke
741
Fluke
743
Fluke
744
Fluke
55XX
Fluke
50
Series
Temperature plays a key role inmany industrial and commercialprocesses. Examples includemonitoring cooking temperature
in food processing, measuringthe temperature of molten steelin a mill, verifying the tempera-ture in a cold storage warehouseor refrigeration system, or regu-lating temperatures in the dryingrooms of a paper manufacturer.
A temperature transmitter willuse a measuring device to sensethe temperature, and then regu-late a 4-20 mA feed-back loop to a controlelement that affectsthe temperature (Fig.1). The control ele-
ment might consist ofa valve that opens orcloses to allow moresteam into a heatingprocess or more fuel
to a burner. The two most com-mon types of temperature sens-ing devices are the thermocouple(TC) and resistive temperature
detector (RTD).Fluke provides a broad rangeof temperature calibration toolsto help you quickly and reliablycalibrate your temperature in-strumentation. A summary of thetemperature calibration capabili-ties of Fluke Process Tools isshown below.
Figure 1.
Measure temperature from an RTD probe
Measure temperature from a T/C probe Simulate an RTD output Simulate an RTD into pulsed excitation current Simulate a T/C output Simultaneous output a T/C, measure mA Simultaneous output an RTD, measure mA Log a temperature measurement Ramp a temperature signal Loop power supply Multifunction Source and Measure Automatically calibrate temperature switches Electronic data capture Upload documented data to PC Integrated HART communication
-
7/31/2019 Calibracion de Temperatura
2/8
2 Fluke Corporation Temperature Calibration
Typical Temperature Calibration Applications
How to calibrate a Thermocouple input transmitter
The Fluke 724 TemperatureCalibrator can provide the threethings necessary to calibrate atemperature transmitter. You cansource a temperature, provideloop power, and measure theresulting output current. Thefollowing example shows how tocalibrate a Type K TC transmitterthat is ranged from 0-150degrees centigrade, generatingan output current range from4-20 mA.
Basic Calibrator Setup
1. Connect the 724 test leads tothe TC transmitter as shown.The output from the thermo-couple jacks on the 724 willsimulate a temperature inputto the transmitter. The redand black test leads willprovide loop power to thetransmitter and will measurethe current resulting fromtemperature changes into thetransmitter.
2. Power on the 724 calibrator.Select the mA button and theLOOP button to select mea-
sure milliamps with 24V looppower applied.
3. Press the Meas/Source buttonuntil the lower portion of the724 display indicates thesource mode.
4. Depress the TC button until aTC type of K is displayed.
5. Select the C button forcentigrade.
6. Set the Zero Point for thisapplication into the Calibra-tor. To do this set the displayinitially to 0.0C. You can usethe up and down arrow keysto change the output value.Use the left and right arrowsto control which decadevalue of the display is beingchanged. When the displayreads 0.0, hold down the 0%key on the 724 and observethat 0% is displayed in thelower right corner of thescreen. This establishes theZero point for calibration.
7. Set the Span Point in theCalibrator. Set the display tothe desired Span value forcalibration. In this examplethe display should read
150C. Depress the 100% keyand observe that 100% isdisplayed in the lower rightcorner of the screen. Thisestablishes the Span point forcalibration.
Performing an As Found Test8. Depress the 0% key; record
the applied temperature andthe corresponding mAmeasurement.
9. Depress the 25% key (2)times; record the appliedtemperature and the corre-sponding mA measurement.
10.Depress the 100% key; recordthe applied temperature andthe corresponding mAmeasurement.
11. Calculate the errors for eachof the (3) points using thefollowing formula: ERROR =([(I-4)/16]-[(T/TSPAN])*100where Error is in % of span, Iis your recorded mA mea-surement, T is the recordedtemperature and TSPAN is thetemperature input span(100% - 0% points). The errorcalculation table belowshows how to apply theformula to actual recordedmeasurements.
12.If your calculated errors areless than the specifiedinstrument tolerance, the
transmitter has passed theAs-Found test. If the test hasnot passed, perform adjust-ments as necessary.
Adjusting the Transmitter13.Depress the 0% key to source
the proper temperature for a4 mA output. Adjust the zeropotentiometer until thecurrent reading is 4.00 mA.
14.Depress the 100% key tosource the proper temperaturefor a 20 mA output. Adjust theSpan potentiometer until thecurrent reading is 20.00 mA.
15.Depress the 0% key againand adjust the zero potenti-ometer again if necessary, toget a 4.00 mA output.
Perform an As Left TestRepeat steps 8 through 12 tocomplete the full calibrationprocedure on your temperaturetransmitter.
mA Measurement TC Source T Span Formula Error %
4.02 0C 150C ([4.02-4)/16]-[0/150])*100 0.1250
11.95 75C 150C ([11.95-4)/16]-[75/150])*100 -0.3125
20.25 150C 150C ([20.25-4)/16]-[150/150])*100 1.5625
Temperature transmitter error calculation example
100%
25%
25%
RECALL
MEAS
SOURCE
STORESETUP
TC RTD
0%
TEMPERATURECALIBRATOR724
V
F
C
V LOOPmA
VmA
MEASURESOURCE/MEASURE
30VMAX ALLTERMINALS
4W
3W V
RTD
TC
COM COM
LOOP
TEST DC PWR++
Rosemount 444
-
7/31/2019 Calibracion de Temperatura
3/8
-
7/31/2019 Calibracion de Temperatura
4/8
4 Fluke Corporation Temperature Calibration
Fluke 725 Multifunction Process Calibrator All the temperature capabilities of Fluke 724
below, plus: Measure pressure with Fluke 700Pxx
Pressure Modules Source or measure frequency to 10 kHz Source mA signals
Fluke 741, 743 and 744 DocumentingProcess Calibrators
Measure or simulate JKTERSBLUNCthermocouples
Measure or simulate 7 RTD types, per 712below, plus Cu 10 (47)
Source or measure volts, ohms, mA, frequency Measure pressure with Fluke 700Pxx
Pressure Modules Electronically capture results of automated
procedures Serial communication to PC (743, 744) HART communication (744)
Fluke 724 TemperatureCalibrator
Measure or simulate JKTERSBLUN thermocouples
Measure or simulate 7 RTDtypes (see 712)
Measure or source VDC, Measure 24 mA, with or
without 24 volt loop supply Source TC or RTD while
measuring V or mA Handles pulsed RTD transmit-
ters with pulses >5 ms
Fluke 712 RTD Calibrator
Measure or simulatePt 100 200 500 1000 (385),Pt 100 (3926), Pt 100 (3916)and Ni 120 (672)
Measure or source 15 to3200
Handles pulsed RTDtransmitters
Selectable F or C Banana jacks for 2W, 3W or
4W measurements
Fluke 714 ThermocoupleCalibrator
Measure or simulate JKTERSBLU thermocouples
Measure or source -10 to 75 mV Selectable F or C Thermocouple mini-jack
termination
-
7/31/2019 Calibracion de Temperatura
5/8
Temperature Calibration Fluke Corporation 5
Fluke 51 Series II and 52 Series II
Large backlit dual display shows any combina-tion of T1, T2 (52 only), T1-T2 (52 only), plusMIN, MAX, or AVG
Relative time clock on MIN, MAX, and AVG pro-vides a time reference for major events.
Electronic Offset function allows compensationof thermocouple errors to maximizeoverall accuracy
Measures J, K, T, and E-types ofthermocouples
Readout in C, F, or Kelvin (K).
Fluke 51 Single InputDigital Thermometer
Fluke 52 Dual InputDigital Thermometer
Fluke 53 Single Input DigitalThermometer with data logging
Fluke 54 Dual Input DigitalThermometer with data logging
The Fluke 53 Series II and 54 Series II
Include all the features listed for the 51and 52 Series II plus: Data Logging up to 500 points of data with user
adjustable recording interval Additional thermocouple types R, S, and N (for a
total of 7 different types) Real time clock captures the exact time of day
when events occur Recall function allows logged data to be
easily reviewed on the meter display IR communication port allows data to be ex-
ported to optional FlukeView Temperature PCsoftware for further analysis and graphing
65 Infrared Thermometer
Range -40 to 500C (-40 to932F)
0.1 resolution up to 200 Measurements in less than
1 second Bright laser beam for easy
targeting 8:1 Optical Resolution MIN/MAX reading capture Large backlit dual display
2620T/2635T RecordingThermometers 20 channel precision
thermometer with SPRTReference Probe
Up to 18 channels TCCalibration System
Up to 9 channels RTDCalibration System
NIST/DKD certifiedcalibration to .05C
61 Infrared Thermometer
Easy to use Quick non-contact
temperature measurementsfrom -18 to 275 C(0 to 525 F)
0.2 C or 0.5 F resolution Backlight illuminates the
display in the dark Laser-sighted; 8:1 optical
resolution
-
7/31/2019 Calibracion de Temperatura
6/8
6 Fluke Corporation Temperature Calibration
Temperature Test Tool Summary Performance: Selected examples
Function Range Resolution Accuracy Notes
RTD Measure -200 to 800C 0.1C 0.3C 8 RTD typesPT100-385 Simulate -200 to 800C
RTD Measure -200 to 800C 0.1C 0.33C 7 RTD typesPT100-385 Simulate -200 to 800C 0.1C 0.33C
Resistance Measure 0-11,000 .01 .05% + 50 m100 range Simulate 0-11,000 .01 .01% + 40 m
Resistance Measure 15-3200 .1 .1 to 1100 range Simulate 15-3200 .1 .1 to 1
Thermocouple Measure -202 to 1374C 0.1C 0.3C JKTERSBLUNCType K Simulate -202 to 1374C 0.1C 0.3C
Thermocouple Measure 0.1C 1000C + 0.3C
Thermocouple Measure 0.1C 1000C + 0.3C
Thermocouple Measure -200 to 1370C 0.1C 0.3C + 10 V JKTERSBLU plusType K Simulate -200 to 1370C 0.1C 0.3C + 10 V N on 724, 725
mV Measure 0-110 mV .001 mV .25% + .015% FSSource 0-110 mV .001 mV .01% + .005% FS
mV Measure -10 to 75 mV .01 mV .025% + 1 countSource -10 to 75 mV .01 mV .025% + 1 count
Measure mA 0-24 mA .001 mA .020% + 2 count Measure mA 0-24 mA .001 mA .010% + .015% FS Loop Power Supply 24V dc n/a 10%
Dry Well Calibrator: A tempera-ture calibrator that uses a
precision oven to source precisetemperature. This style ofcalibrator is often used for theverification of temperaturesensors.
Excitation current: A constantcurrent applied to an RTD probeto determine actual resistancefor temperature measurement.Typical values are 2 mA or lessto minimize self-heating of theprobe.
IPTS-68: International Practical
Temperature Scale of 1968. Atemperature standard adopted in1968 that uses intrinsic stan-dards to define the measurementof temperature.
ITS-90: International Tempera-ture Scale of 1990. A tempera-
ture standard adopted in 1990that uses intrinsic standards todefine the measurement oftemperature. This standardmodifies the intrinsic standardsof IPTS-68 with additionalintrinsic references.
Lead Resistance Compensation:A compensation method usedwith 3 and 4 wire RTDs andresistance measurement. Thismethod negates the error associ-ated with lead resistance whenmaking an RTD measurement.
Reference Temperature: Thetemperature at which a thermo-couple temperature measure-ment is referenced. For FlukeCalibration tools, this is thetemperature where the thermo-couple mini-plug is connected tothe calibrator.
R: The resistance value of anRTD probe at 0C. Example
PT100-385, R = 100.RTD: Resistance TemperatureDevice, a temperature measure-ment sensor that has predictablechanges in resistance with achange in temperature. The mostcommon RTD is the platinumPT100-385.
Seebeck Effect: Thermoelectriceffect in which the voltagepotential increases with tem-perature (thermocouples) in ajunction of dissimilar metals.
Triple point of water: Thistemperature reference point isthe intrinsic standard at whichwater is liquid, ice and gas. Thisreference point defines 0.01C.
Temperature Terminology
744
743/743B
741/741B
725
724
714
712
53/54
51/52
-
7/31/2019 Calibracion de Temperatura
7/8
Temperature Calibration Fluke Corporation 7
Alloys Fluke 700 Series DPCs
Plug + - Temperature EMF RangeColor Range
White Pt-30% Rh Pt-6% Rh 600 to 1820C, 1.792 to 13.820 mV
B Platinum- Platinum- 1112 to 3308F30% Rhodium 6% RhodiumRed W-5% Re W-26% Re 0 to 2316C, 0 to 37.079 mV
C Tungsten- Tungsten- 32 to 4201F5% Rhenium 26% RheniumPurple Ni-Cr Cu-Ni -250 to 1000C, -9.719 to 76.370 mV
E Nickel- Copper-Nickel -418 to 1832FChromium (Constantan)(Chromel)
Black Iron Fe Cu-Ni -210 to 1200C, -8.096 to 69.555 mV
J Copper-Nickel -346 to 2193F(Constantan)Yellow Ni-Cr Ni-Al -200 to 1372C, -5.891 to 54.886 mV
K Nickel- Nickel- -328 to 2502FChromium Aluminum(Chromel)Blue Fe Cu-Ni -200 to 900C, -8.166 to 53.147 mV
L Iron Copper-Nickel -328 to 1652F(Constantan)Orange Ni-Cr-Si Ni-Si-Mg -200 to 1300C, -3.990 to 47.514 mV
N Nickel-Chromium- Nickel-Silicon- -328 to 2372FSilicon Magnesium(Nicrosil) (Nisil)
Green Pt-13% Rh Pt -20 to 1767C, -0.101 to 21.089 mV
R Platinum- Platinum -4 to 3213F13% RhodiumGreen Pt-10% Rh Pt -20 to 1767C, -0.103 to 18.682 mV
S
Platinum- Platinum -4 to 3213F10% Rhodium
Blue Cu Cu-Ni -250 to 400C, -6.181 to 20.873 mV
T Copper Copper-Nickel -418 to 752F(Constantan)White Cu Cu-Ni -200 to 600C, -5.693 to 34.320 mV
U Copper Copper-Nickel -328 to 1112F
Thermocouples TemperatureConversion
F
4352
4172
3992
3812
3632
3452
3272
3092
2912
2732
2552
2372
2192
2012
1832
1652
1472
1292
1112
932
752
572
392
212
32
-148
-328
C
2400
2300
2200
2100
2000
1900
1800
1700
1600
1500
1400
1300
1200
1100
1000
900
800
700
600
500
400
300
200
100
0
-100
-200
-459 -273
Fixed Points of ITS-90
961.78C Silver freezing point
660.323C Al freezing point
419.527C Zinc freezing point
-38.8344C Mercury triple point
231.928C Tin freezing point
156.5985C Indium freezing point
0.01C H2O triple point
29.7646C Gallium melting point
-189.3442C Argon triple point-218.7916C Oxygen triple point-248.5939C Neon triple point-259.3467C Hydrogen triple point
Application AssistanceTo request copies of these application notes,contact your Fluke sales organizations.
Understanding Specifications (pub ID 10127) Transmitter Calibration (pub ID 10123) HART Transmitter Calibration (pub ID 10179) Limit Switches (pub ID 10121) Pressure Modules (pub ID 10188) Loop Calibration (pub ID 10437)
-
7/31/2019 Calibracion de Temperatura
8/8
Fluke 700TC1 ThermocoupleMini-Plug Kit (11 types)For use with: Fluke 700, 720 or
740 Series Process Calibrators, Fluke714 Thermocouple Thermometer.Description: A kit of 10 mini-plugconnectors:Type J (black), oneType K (yellow), oneType T (blue), oneType E (purple), oneType R/S (green), oneType B or Cu (white), oneType L (J-DIN) (blue), oneType U (T-DIN) (brown), oneType C (red), oneType N (orange), one
Fluke 700TC2 ThermocoupleMini-Plug Kit Type J, K, T, E, R/SFor use with: Fluke 700, 720 or 740Series Process Calibrators, Fluke 714Thermocouple Thermometer.Description: A kit of seven mini-plugconnectors:Type J (black), twoType K (yellow), twoType T (blue), oneType E (purple), oneType R/S (green), one
80T-IR Non-Contact InfraredTemperature ProbesFor quick, non-contact temperaturemeasurement in less than onesecond Instantly measures
temperatures from -18 C to260 C (0 to 500 F)
Highly repeatable readingsdisplayed in C or F
Optical resolution 4:1 Output 1mV/C or 1mV/F
(switch selectable)
80T-IR/E Extended Range,Non-Contact InfraredTemperature ProbeSpecial version of the 80T-IR:
Extended range: 0 to 1,000 F Measures F only (no C)80TK Thermocouple ModuleConverts multimeters with a mVrange into digital thermometers. Converts K-type thermocouple
signals into mV output. Connects to DMM via
standard bannana plugs Switch selectable C or F Range: -50 C to 1000 C
(-58 to 1832 F)
Infrared Temperature Probe -80PK-IRCompatible with Fluke 714, 724,
725, 74X, Fluke 50 Series thermom-eters, Fluke 16 and 89 IV DMMs, thisprobe measures the temperature ofobjects that cant be touched. Measurement range: -18 to
260 C (0 to 500 F)
General Purpose Bead Probes80PK-1 and 80PJ-1Basic, inexpensive thermocouple.Good accuracy and fast response.(Not suitable for liquid immersion.) 80PK-1 for K-type thermometers;
80PJ-1 works for J-Typethermometers
Measurement range: -40 to260 C (-40 to 500 F)K-Type Immersion Probe -80PK-2AGeneral-purpose probe includingliquids and gels. Measurement range: -40 to
982 C (-40 to 1800 F)
K-Type Flat Surface Probe -80PK-3AMeasure the temperature of flat orslightly convex surfaces, such as hotrollers and plates. Exposed junction for direct
contact with surface Measurement range: 0 to 260 C(32 to 500 F)
K-Type Air Probe80PK-4AFor air and non-caustic gasapplications. Probe of 304 stainless steel with
perforated baffle around the bead Measurement range: -40 to 816C
(-40 to 1500 F)
Piercing Probes -80PK-5A and 80PT-5ADesigned for foods, liquids and gels.
Probe made of 304 stainless steel 80PK-5A is compatible withK-Type instruments; 80PT-5Ais designed for use with T-Typethermometers
Measurement range: -40 to260 C (-40 to 500 F)
K-type Exposed Junction Probe -80PK-6AUse both as a surface probe and forinsertion into air and non-corrosivegases. Probe made of 304 stainless steel Measurement range: -40 to 816 C
(-40 to 1500 F)
K-Type Industrial / Heavy DutySurface - Probe - 80PK-7For flat or slightly curved surfaces.
Exposed junction allows directcontact with the surface beingmeasured
Rugged design for extendedservice life
Measurement range: -127 to600 C (-196 F to 1112 F)
K-Type Probe Pipe Clamp Tem-perature Probe-80PK-8Designed for pipes from 6.4 mm(1/4") diameter to 34.9 mm(1-3/8") diameter. Measurement range: -29 C to
149C (-20 F to 300 F)
General Purpose Probes -80PK-9 and 80PJ-9Use as surface probe and forinsertion into air and non-causticgases. Probe constructed of 304
stainless steel 80PK-9 is compatible with K-type
thermometers; 80PJ-9 operateswith J-Type
Measurement range:-40 to 260 C (-40 to 500 F)
Extension Wire Kits - 80PK-EXT,80PJ-EXT and 80PT-EXT
For extending and repairing J, K orT-type thermocouple wires. Kit includes 3 meters of thermo-
couple wire and 1 pair ofmale/female mini-connectors
Maximum continuous exposuretemperature: 260 C (500 F)
Temperature Accessories
Fluke CorporationPO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.PO Box 1186, 5602 BDEindhoven, The Netherlands
For more information call:U.S.A. (800) 443-5853 orFax (425) 446-5116Europe/M-East/Africa(31 40) 2 675 200 orFax (31 40) 2 675 222Canada (800) 36-FLUKE orFax (905) 890-6866
Other countries +1 (425) 446-5500 orFax +1 (425) 446-5116Web access: http://www.fluke.com
2003 Fluke Corporation. All rights reserved.Printed in the Netherlands 4/2003 Pub ID 10308-engRev 02
Fluke. Keeping your worldup and running.